Build overview
Planning your BrainSaw build
This is a high-level overview of the steps required to assemble a BakingTray-based STPT system. Other pages on the site provide more detail. There is no complete end-to-end set of build instructions, but groups have successfully used what is on here to autonomously build a BrainSaw.
Step 1: Decide on your build path
If you have good general tech skills and enjoy building stuff, you can probably build a BrainSaw STPT system. There are several paths to assembling a system, and there is likely one to fit your skill-set.
Modify an existing two-photon microscope
Upgrade a legacy TissueCyte 1000 system from TissueVision
Buy a suitable new two-photon microscope from a trusted vendor and add serial section capability to that
DIY the entire build yourself from scratch
The SWC AMF can provide advice and potentially even hands-on assistance for your build. Contact Rob Campbell at the SWC AMF to register your interest.
Step 2: Gather the parts
The components can be collated and tested in parallel.
Put in place a suitable acquisition PC with ScanImage.
Put in place the two-photon microscope, ensuring space for the stages and slicing hardware.
Quantify microscope performance, measuring FoV, field homogeneity, PSF, laser power, etc. See also MPQC.
Assemble the vibratome and associated hardware. Test these.
Assemble the X/Y/Z stage and test.
Make the water bath.
Set up an analysis PC for stitching with StitchIt.
For further information on parts see hardware requirements.
Step 3: Assemble
Get BakingTray communicating with the stages, laser, and vibratome, then fill in the settings files.
Use MPQC and BakingTray to calibrate laser power in ScanImage at the wavelengths you plan to use often. We normally calibrate at 770, 800, 920, 950, and 1040 nm.
Clamp the water bath into the X/Y/Z stage. Set the X/Y stages to their mid-points (0 mm) and centre the frosted part of a slide under the objective.
Set the X, Y, Z limits of the stages in BakingTray so the objective can not hit any part of the bath.
Level the stage platform using a thin fluorescent slide.
Ensure the X and Y stage axes move the sample along the scan axes.
Calibrate the number of microns per pixel and set up the image sizes in
frameSizes.yml.Position the vibratome, add the blade holder, and update the stage limits.
Use fluorescent agarose to adjust blade angle. This is also a chance to test slicing.
Enter the blade and slide positions into the system settings file.
Run test samples and fine-tune stitching accuracy.